Urine Analyzer NUNZ-200 handles 240 tests per hour for chemistry and 50 to 120 tests per hour for microscopy and combined modes. It has a sample loading capacity of 60 for uninterrupted processing. This analyzer includes a built-in barcode reader to ensure fast and accurate sample identification. It features a self-learning function that enhances recognition accuracy with continued use. Our unit uses three counting chambers to deliver precise and consistent analysis.
$29,575.00
Specifications
| Throughput | Chemistry: 240T/H Microscopy: 50 to 120T/H Chem + Micro: 50 to 120T/H |
| Min. sample volume | 5ml |
| Sample preparation | Native sample |
| Counting chamber | 3 |
| Detection rate | ≥98% for the sample at a concentration of 5 cells/μl |
| Accuracy | For chemistry measurement, compared to the reference solution, results are within tolerance of one level, or no false positives or negatives For microscopy measurement, ≥95% |
| Repeatability | For chemistry measurement, CV≤1% For microscopy measurement, CV≤7% |
| Carryover | At a concentration of 4600 to 5400 cells/μl, ≤ 1 cells/μl At a concentration of 9200 to 10800 cells/μl,≤ 2 cells/μl |
| Verification | Full-view field images + segmented images |
| Report | Comprehensive results with images and diagrams |
| Sampler | Loading capacity of 60 samples |
| Strip tank capacity | 200 strips |
| STAT | Available |
| Data interface | Bidirectional |
| LIS/HIS connection | Available |
| Computer | External |
| Results storage | ≥200,000 results |
| Built-in barcode reader | Yes |
| Packaging dimension | 825×647.5×514mm |
| Weight | 75 kg |
Features
Applications
Standard Accessories
Tube rack-1pc
Power cord-1 pc
Optional Consumables
1. Calibrator for Urine Formed Elements Analyzer
High Concentration 28×10m
Medium Concentration 28×10ml
Low Concentration 28×10ml
Sensitivity 28×10ml
Mixed Pack (7×H, 7×M, 7×L,7×S)
2. Urine Chemistry Quality Control
Mixed Pack (10ml×4×3)
3. Urine Test Strip (100 strips/box)
4.Reagents
Cleanser II (40ml concentrate for 400ml)
Buffer (40ml concentrate for 400ml)
Reagent (cleanser I:1×500ml, cleanser II: 1×40ml, Buffer: 1×40ml)
Ask Expert
FAQ for Urine Analyzer NUNZ-200
1: How accurate is the Urine Analyzer NUNZ-200 in detecting microscopic particles?
Our Urine Analyzer offers a high detection rate of greater than or equal to 98% for particles at a concentration of 5 cells/ microliter. The system also provides precise RBC morphological analysis and follows ISLH reference standards for consistent results.
2: Does the Urine Analyzer support visual verification of results?
Yes, the Urine Analyzer provides full-view field images and segmented images for each test. This allows on-screen verification and confirmation without needing to repeat measurements.
3: How much sample volume is required for each test?
The Urine Analyzer requires a minimum of 5 mL of urine per test. This small sample volume makes it suitable even when limited patient samples are available.
4: What parts of the Urine Analyzer NUNZ-200 need regular cleaning?
The sample chamber, strip feeder, counting chambers, and waste areas should be cleaned regularly. Cleaning these components ensures precise measurements and prolongs the life of the analyzer.
5: How often does the Urine Analyzer NUNZ-200 require maintenance?
Routine maintenance is minimal and typically involves daily cleaning of the sample path and checking consumables. Periodic system calibration and software updates ensure long-term accuracy and performance.