Automated Solid Phase Extractor NSPE-102 supports four-channel extraction at 120 mL/min with nitrogen purging. This system uses waste segregation to reduce contamination between parallel extractions. Our elevated frame improves visibility across the platform and offers easier access during sample loading. It adjusts flow automatically to limit manual handling during operations. This extractor features illuminated status bars that provide real-time visual process updates and alerts.
| Channels | 4 |
| Sample Volume | 0.1 mL to 20 mL |
| Sample Tube Volume | 10 to 20 mL, customizable |
| Column Pressure | 0 to 145 psi |
| Receiving Volume | 10 mL / 15 mL, other customized |
| Channel Independent Control | Support |
| Pump Flow Rate | 0.1 to 120 mL/min |
| Sample Pump | Precision Continuous Sample Pump |
| Types of Solvents | 8-seed, 500 mL Solvent Bottle |
| Puncture Sampling | Supporting |
| Loading Mode | X/Y/Z Three-Dimensional Manipulator |
| Accuracy of Manipulator | 0.01 mm |
| Sealing Technology | Docking Seal & Pan-plug Seal |
| Extraction Mode | Positive Pressure Extraction |
| Pressure Sensing | Support, Overvoltage Alarm, Editable (5 to 145 psi) |
| Column Drying Method | Nitrogen Purge |
| Waste Collection | Branch collection, 3 channels |
| A/B in Series | Support, A/B Pillar |
| Cleaning Mode | Nitrogen, solvent, and stirring cleaning |
| Liquid Level Following | Support |
| Solvent Residual Capacity | Solvent deficiency hints |
| Working Temperature | 0 to 40°C |
| Working Humidity | 10 to 85% |
| Power Consumption | 150 W |
| Power Supply | 220/110V, 50/60Hz |
| Dimensions (H×W×D) | 680×580×670 mm |
Automated Solid Phase Extractor NSPE‑102 enhances analytical throughput by processing four samples simultaneously. It is suited for central diagnostics, toxicology, and environmental monitoring laboratories.