Pulse Vacuum Autoclave NPVA-104 provides a 2500L capacity and operates with an inner chamber pressure range of -0.1 to 0.3 Mpa. It features a rectangular chamber and a durable ring-type jacket for reliable operations. This unit includes an independent steam inlet system to prevent contamination. It uses a high-performance filter for improved cleanliness and sterilization effectiveness. Our pulse vacuum autoclave is controlled by a PLC and a color touch screen for easy operation.
| Capacity | 2500L |
| Pressure of inner chamber | -0.1 to 0.3Mpa |
| Pressure of jacket | 0.3Mpa |
| Designed temperature | 150°C |
| Working temperature | 139°C |
| Working pressure | 0.25Mpa |
| Vacuum leakage rate | ≤0.13KPa/min |
| Ultimate vacuum | -97Kpa |
| Opening pressure for the inner chamber’s safety valve | 0.28Mpa |
| Opening pressure for jacket’s safety valve | 0.28Mpa |
| Precision of sterilization temperature control | 0 to 2°C |
| Temperature uniformity | ±1°C |
| Negative pressure pulsation vacuum amplitude and times | Amplitude -80 to 0Kpa, Times 1 to 99 |
| Times of pulses across pressure | Amplitude -80 to 80Kpa, Times 1 |
| Times of positive pressure pulses | 1 to 3 |
| Steam consumption | 132Kg/C |
| Tap water consumption | 1250Kg/C |
| Power consumption | 380V, 50Hz |
| Power supply | 6kW |
| Internal dimension | 900×1380×2000mm |
| External dimension | 2360×1860×2230mm |
| Packaging dimension | 2350×1930×2480mm |
| Weight | 2700Kg |
Pulse Vacuum Autoclave NPVA-104 is used to enhance sterilization efficiency by removing air from materials, allowing for better penetration of steam. It is commonly applied in medical, pharmaceutical, and laboratory settings.