WhatsApp

Gasoline Oxidation Stability Tester NGOS-100 Catalog

Overview

Gasoline Oxidation Stability Tester NGOS-100 comes with a 28.8 L capacity and maintains temperature control with ±0.1°C accuracy. It features automatic leak detection to reduce manual workload. This unit has a window-based interface display for easy control and monitoring. It supports dual-group testing with an automated process and data printing. Our tester includes a large hyperbolic touch screen for clear and intuitive operation.

Specifications:

Capacity 28.8 L
Temperature range 98 to 102°C
Temperature control accuracy ±0.1°C
Temperature control mode PID control, Automatic control
Temperature measuring element Platinum resistance (PT100)
Thermometer graduation 0.1°C
Pressure gauge 0 to 1.6 MPa, 0.4 level
Bath depth 320 mm
Stirring method Electric
Standards ASTM D525, ISO7536, DIN51780, IP40
Power consumption 1500W
Power supply AC220V 50Hz
Dimensions 630 × 560 × 490 mm
Packing dimensions 830 × 820 × 850 mm
Net weight 42 kg
Gross weight 52 kg
Features:
  • > Stainless steel connections
  • > Durable construction
  • > Stable control system
  • > Dual-channel display
  • > Fault alarm function
Applications:

Gasoline Oxidation Stability Tester NGOS-100 is used in petroleum refineries, fuel research labs, and quality control departments to assess the oxidation stability of gasoline.

Standard Accessories:

Oxygen Bomb × 2 pcs
Wrench × 1 pc
Safety Pin × 4 pcs
Oxygen Tube × 1 pc
Glass Sample Bottle × 2 pcs
Thermometer × 1 pc